Quickie QM-7 Series Manuel de service

Naviguer en ligne ou télécharger Manuel de service pour Non Quickie QM-7 Series. Quickie QM-7 Series Service manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - 120938 Rev A

Quickie QM-7 Series Service Manual©2012 Sunrise Medical Inc. 120938 Rev A

Page 2

QM-7 SERIES SERVICE MANUALPAGE 1.6JANUARY 2012 SUNRISE MEDICAL1. Hand-Held Transceivers: Antenna is usually mounted directly on the unit.These incl

Page 3 - Introduction

QM-7 SERIES SERVICE MANUALPAGE 1.7 JANUARY 2012SUNRISE MEDICALTo help prevent Electro Static Discharge (ESD) the following proper handling technique

Page 4 - Basic Setup

QM-7 SERIES SERVICE MANUALPAGE 1.8JANUARY 2012 SUNRISE MEDICAL• Batteries function as a power wheelchair’s fuel tank. The larger the group size, t

Page 5

QM-7 SERIES SERVICE MANUALPAGE 1.9 JANUARY 2012SUNRISE MEDICALVR2 Controller ButtonsBattery Gauge A series of ten LED’s, which indicate charge level

Page 6

QM-7 SERIES SERVICE MANUALPAGE 1.10JANUARY 2012 SUNRISE MEDICAL1 = 24 Vdc (POS)2 = 0 Vdc (NEG)3 = Inhibit 1/ Programmer1 = White2 = Blue3 = Bl

Page 7

QM-7 SERIES SERVICE MANUALPAGE 1.11 JANUARY 2012SUNRISE MEDICALOn/Off Key- Press to power on or off the power chair or Controller.Horn Key- Activa

Page 8

QM-7 SERIES SERVICE MANUALPAGE 1.12JANUARY 2012 SUNRISE MEDICALOn/Off Key- Press to power on or off the power chair or Controller.Speed down butto

Page 9

QM-7 SERIES SERVICE MANUALPAGE 1.13 JANUARY 2012SUNRISE MEDICALM1 = LEFT SIDE MOTORM2 = RIGHT SIDE MOTORINH-2 = INHIBIT 2A1 = ACTUATOR 1A2 =(NOT USE

Page 10 - Section 1

QM-7 SERIES SERVICE MANUALPAGE 1.14JANUARY 2012 SUNRISE MEDICALQM-7 Series ModulesSection 1OMNI (BOTTOM)User Jack 2 User Jack 1Port 2 “D” Type Port

Page 11

QM-7 SERIES SERVICE MANUALPAGE 2.1 JANUARY 2012SUNRISE MEDICAL This list of tools will be needed to accomplish all of the tasks given in this Te

Page 12

QM-7 SERIES SERVICE MANUALPAGE 2JANUARY 2012 SUNRISE MEDICALQuickie QM-7 Series Service Manual ContentsCIntroduction and Basic SetupMotor - Drive G

Page 13

QM-7 SERIES SERVICE MANUALPAGE 2.2JANUARY 2012 SUNRISE MEDICALSection 2 Power Module BUS Connections - VR2Power ModuleVR2-904-WayTyco BusBASESEATMB

Page 14

QM-7 SERIES SERVICE MANUALPAGE 2.3 JANUARY 2012SUNRISE MEDICALSection 2 DriveRightDriveLeftPower ModuleR-Net EL-904-WayIntechMotor4-WayIntechMotor2-

Page 15

QM-7 SERIES SERVICE MANUALPAGE 2.4JANUARY 2012 SUNRISE MEDICALDual-ToggleSingle ActuatorDriver4-Way AmpMate-N-Loc6-WayMini-Fit J r.Buddy ButtonSwit

Page 16

QM-7 SERIES SERVICE MANUALPAGE 2.5 JANUARY 2012SUNRISE MEDICAL3-AXIS ACTUATOR DRIVER - TILTSwitch ControlsACTUATORHARNESSACTUATORHARNESSSwitchSeat T

Page 17

QM-7 SERIES SERVICE MANUALPAGE 2.6JANUARY 2012 SUNRISE MEDICAL 2-AXIS ACTUATOR DRIVERSwitch ControlsACTUATORHARNESSSwitch6-WayMini-Fit Jr.Seat Lift

Page 18

QM-7 SERIES SERVICE MANUALPAGE 2.7 JANUARY 2012SUNRISE MEDICAL R-NET MULTI-ACTUATORACTUATORHARNESSACTUATORHARNESSACTUATORHARNESSIntellegentSeatingMo

Page 19 - Section 2

QM-7 SERIES SERVICE MANUALPAGE 2.8JANUARY 2012 SUNRISE MEDICAL 4-Way PGDBusAttendant Contr olConnector Block4 X CAN4-WayPGD Bus4-WayPGD Bus4-WayPGD

Page 20

QM-7 SERIES SERVICE MANUALPAGE 3.1 JANUARY 2012SUNRISE MEDICALSection 3Batteries should last an average of 1 to 1.5 years.Listed below are some Fact

Page 21

QM-7 SERIES SERVICE MANUALPAGE 3.2JANUARY 2012 SUNRISE MEDICALSection 3Battery Connection Test1. Check that the female VR2 Bus plug on the chair h

Page 22

QM-7 SERIES SERVICE MANUALPAGE 3.3 JANUARY 2012SUNRISE MEDICALSection 3VR2 Remote Controller DisplayMaximum Speed Indicator Ripples. This Indicates

Page 23 - 3-AXIS ACTUATOR DRIVER - LIFT

QM-7 SERIES SERVICE MANUALPAGE 0.1 JANUARY 2012SUNRISE MEDICALPlease read and follow instructions in this service manual before attempting to troubl

Page 24

QM-7 SERIES SERVICE MANUALPAGE 3.4JANUARY 2012 SUNRISE MEDICALTroubleshooting with VR2 Diagnostic CodesSection 31. Check that the batteries are ful

Page 25

QM-7 SERIES SERVICE MANUALPAGE 3.5 JANUARY 2012SUNRISE MEDICAL1. Check that the batteries are fully charged and in good condition; and check all ca

Page 26

QM-7 SERIES SERVICE MANUALPAGE 3.6JANUARY 2012 SUNRISE MEDICALSection 33. Otherwise, check the brushes on the Right motor (figure 3.6.1). Using a

Page 27 - Section 3

QM-7 SERIES SERVICE MANUALPAGE 3.7 JANUARY 2012SUNRISE MEDICALSection 3A joystick trip is indicated. Make sure that the joystick is in the center po

Page 28

QM-7 SERIES SERVICE MANUALPAGE 3.8JANUARY 2012 SUNRISE MEDICALSection 3Nine Bars - Solenoid Brake TripCheck that the batteries are fully charged, a

Page 29

QM-7 SERIES SERVICE MANUALPAGE 4.1 JANUARY 2012SUNRISE MEDICALSection 4R-net TroubleshootingChair Will Not Power Up2131. Check for battery voltage a

Page 30

QM-7 SERIES SERVICE MANUALPAGE 4.2JANUARY 2012 SUNRISE MEDICALSection 44. Verify that the bus cables are correctly mated between the hand control

Page 31

QM-7 SERIES SERVICE MANUALPAGE 4.3 JANUARY 2012SUNRISE MEDICALSection 46. Verify that battery voltage is present at each connector leading to the b

Page 32

QM-7 SERIES SERVICE MANUALPAGE 4.4JANUARY 2012 SUNRISE MEDICALR-net Fault CodesSection 4Fault Codes for Hand Control or OmniThe R-Net control used

Page 33

QM-7 SERIES SERVICE MANUALPAGE 4.5 JANUARY 2012SUNRISE MEDICALSection 4Trip Text Corrective ActionOver-current This fault occurs when the limits of

Page 34

QM-7 SERIES SERVICE MANUALPAGE 0.2JANUARY 2012 SUNRISE MEDICALWhen setting up the components of the chair, complete the following checklist to ensu

Page 35 - Section 4

QM-7 SERIES SERVICE MANUALPAGE 4.6JANUARY 2012 SUNRISE MEDICALSection 4Note: On hand controls that contain LED battery gauges, they will display t

Page 36

QM-7 SERIES SERVICE MANUALPAGE 4.7 JANUARY 2012SUNRISE MEDICALSection 4Example of R-10 Fault Isolation1. In this example, the right motor has been

Page 37 - Cable Tie

QM-7 SERIES SERVICE MANUALPAGE 4.8JANUARY 2012 SUNRISE MEDICALSection 43. Locate the creep micro-switch mounted at the rear of the actuator (A)

Page 38

QM-7 SERIES SERVICE MANUALPAGE 4.9 JANUARY 2012SUNRISE MEDICALSection 4Chair Won't Drive in Creep Speed When Tilted An external micro-switch is

Page 39

QM-7 SERIES SERVICE MANUALPAGE 4.10JANUARY 2012 SUNRISE MEDICALSection 4Tilt/Lift Will Not Operate1. Select tilt on the hand control (fi gure. 4.10

Page 40

QM-7 SERIES SERVICE MANUALPAGE 4.11 JANUARY 2012SUNRISE MEDICALSection 4Warning: Use caution in the next step. Do not short the leads of the Multi

Page 41

QM-7 SERIES SERVICE MANUALPAGE 4.12JANUARY 2012 SUNRISE MEDICALSection 4 Positive probe hereNegative probe hereContinuity between pinsPositive pro

Page 42

QM-7 SERIES SERVICE MANUALPAGE 5.1 JANUARY 2012SUNRISE MEDICALSection 5Battery Testing Note: The QM-710 power chair uses either (2) 22NF, or GRP 24

Page 43

QM-7 SERIES SERVICE MANUALPAGE 5.2JANUARY 2012 SUNRISE MEDICALSection 5d. Operate the power chair at high speed by operating the joystick in a hig

Page 44

QM-7 SERIES SERVICE MANUALPAGE 5.3 JANUARY 2012SUNRISE MEDICALSection 53. The individual voltage of each battery may also be checked using a Multim

Page 45

QM-7 SERIES SERVICE MANUALPAGE 1.1 JANUARY 2012SUNRISE MEDICALFor Powerchairs, the multimeter is one of the most useful tools in the toolbox (fi gure

Page 46 - *SEE NOTE*

QM-7 SERIES SERVICE MANUALPAGE 6.1 JANUARY 2012SUNRISE MEDICALNote: This procedure may be done with or without removing the motor from the power ch

Page 47 - Section 5

QM-7 SERIES SERVICE MANUALPAGE 6.2JANUARY 2012 SUNRISE MEDICAL6. Inspect the brush assembly (fi gure. 6.2.1). Replace any brush assembly less than

Page 48

QM-7 SERIES SERVICE MANUALPAGE 6.3 JANUARY 2012SUNRISE MEDICALSection 69. Measure the resistance of the park brake assembly by connecting a Multimet

Page 49

QM-7 SERIES SERVICE MANUALPAGE 7.1 JANUARY 2012SUNRISE MEDICALBattery Access/Removal 2. Remove cushion and Pivot the entire Seat/Back frame assembl

Page 50 - Section 6

QM-7 SERIES SERVICE MANUALPAGE 7.2JANUARY 2012 SUNRISE MEDICALBatteries installed incorrectly can blow the fuse for this system. Pay careful atten

Page 51

QM-7 SERIES SERVICE MANUALPAGE 7.3 JANUARY 2012SUNRISE MEDICALSection 7Motor and Gearbox Removal 1. The Motor must be disconnected from power before

Page 52

QM-7 SERIES SERVICE MANUALPAGE 7.4JANUARY 2012 SUNRISE MEDICALSection 7Removal for units with Mechanical Brake 1. The process for units with a Mech

Page 53 - Section 7

QM-7 SERIES SERVICE MANUALPAGE 7.5 JANUARY 2012SUNRISE MEDICALSection 73. Use 4 mm Hex Key to remove left and right screws/washers attaching the co

Page 54

QM-7 SERIES SERVICE MANUALPAGE 7.6JANUARY 2012 SUNRISE MEDICALSection 71. To remove or replace the Caster Stem, Use a Flat Head Screwdriver to rem

Page 55

QM-7 SERIES SERVICE MANUALPAGE 7.7 JANUARY 2012SUNRISE MEDICALSection 7Gas Spring Removal 2. For Gas Spring Removal, Remove the Lower holding bolt

Page 56

QM-7 SERIES SERVICE MANUALPAGE 1.2JANUARY 2012 SUNRISE MEDICALSection 1AC - ALTERNATING CURRENT - Use this when you want to test something that has

Page 57

QM-7 SERIES SERVICE MANUALPAGE 7.8JANUARY 2012 SUNRISE MEDICAL7. To re-install the Gas Spring, Reverse the previous procedures. Be sure to adjust

Page 58

QM-7 SERIES SERVICE MANUALPAGE 7.9 JANUARY 2012SUNRISE MEDICALSection 71. Remove the seating system to expose the actuator. Disconnect the 6-pin con

Page 59

QM-7 SERIES SERVICE MANUALPAGE 7.10JANUARY 2012 SUNRISE MEDICALSection 74. Remove the actuator and micro-switch assembly as one part if the entire

Page 60

QM-7 SERIES SERVICE MANUALPAGE 7.11 JANUARY 2012SUNRISE MEDICALSection 73. Use a 9/64" Hex wrench to remove the socket head cap screws that at

Page 61

QM-7 SERIES SERVICE MANUALPAGE 7.12JANUARY 2012 SUNRISE MEDICALSection 7Note: You do not have to fully disconnect the gas strut. Once the top moun

Page 62

QM-7 SERIES SERVICE MANUALPAGE 7.13 JANUARY 2012SUNRISE MEDICALSection 78. Use a 1/4" Hex wrench and a 9/16" open-end wrench and remove t

Page 63

QM-7 SERIES SERVICE MANUALPAGE 7.14JANUARY 2012 SUNRISE MEDICALSection 710. Unplug all of the connectors to the recline actuator (fi gure 7.14.2) 1

Page 64

QM-7 SERIES SERVICE MANUALPAGE 7.15 JANUARY 2012SUNRISE MEDICALSection 7Elevate Actuator Removal1. Locate the ISM control box on the back of the ch

Page 65

QM-7 SERIES SERVICE MANUALPAGE 7.16JANUARY 2012 SUNRISE MEDICALSection 72. To remove the actuator, (fi gure 7.16.1) there are 2 sets of attaching h

Page 66

QM-7 SERIES SERVICE MANUALPAGE 8.1 JANUARY 2012SUNRISE MEDICALSection 8ASAP II Seating 16” - 20” wide is shown.The ASAP II seating system is fully

Page 67

QM-7 SERIES SERVICE MANUALPAGE 1.3 JANUARY 2012SUNRISE MEDICALHealth and SafetyWhile working on powered mobility products, it is essential to observ

Page 68

QM-7 SERIES SERVICE MANUALPAGE 8.2JANUARY 2012 SUNRISE MEDICALSection 86. Loosen the Back bracket on the chair to allow for changes to the seat wid

Page 69 - Section 8

QM-7 SERIES SERVICE MANUALPAGE 8.3 JANUARY 2012SUNRISE MEDICALSection 88. After checking alignment, symmetry, and measurement of the final adjustme

Page 70

QM-7 SERIES SERVICE MANUALPAGE 8.4JANUARY 2012 SUNRISE MEDICALSection 8fig. 8.4.14. Once desired height is achieved, Reset placement of bolts then

Page 71

QM-7 SERIES SERVICE MANUALPAGE 8.5 JANUARY 2012SUNRISE MEDICALSection 84. Left Command (4) - Operates extension of leg outward and inward. Actuato

Page 72

QM-7 SERIES SERVICE MANUALPAGE 8.6JANUARY 2012 SUNRISE MEDICALSection 8fig. 8.6.13. To make a small angle change to one of four possible settings,

Page 73

QM-7 SERIES SERVICE MANUALPAGE 8.7 JANUARY 2012SUNRISE MEDICALSection 8Rear Bracket Seat height adjustment1 The adjustment can be done by removing

Page 74

QM-7 SERIES SERVICE MANUALPAGE 8.8JANUARY 2012 SUNRISE MEDICALSection 83. Re-insert and tighten the bolts securing the pivot point. (figure 8.8.1)

Page 75

QM-7 SERIES SERVICE MANUALPAGE 8.9 JANUARY 2012SUNRISE MEDICALSection 8Cantilevered Armrest adjustments1. The Kids Cantilevered Armrest angle adjust

Page 76

QM-7 SERIES SERVICE MANUALPAGE 8.10JANUARY 2012 SUNRISE MEDICALSection 83. For Extreme angle adjustments, remove the outside shell of the armrest c

Page 77

QM-7 SERIES SERVICE MANUALPAGE 9.1 JANUARY 2012SUNRISE MEDICALSection 9Reference and Options Non-expandable 3-Axis controls1. 2-Axis box 4.5 and 7.

Page 78

QM-7 SERIES SERVICE MANUALPAGE 1.4JANUARY 2012 SUNRISE MEDICALIf mishandled batteries can be dangerous and hazardous.• All mobility batteries, whe

Page 79 - Section 9

QM-7 SERIES SERVICE MANUALPAGE 9.2JANUARY 2012 SUNRISE MEDICALSection 9Powered elevating or articulating legrestThe actuator buttons can operate an

Page 80

QM-7 SERIES SERVICE MANUALPAGE 9.3 JANUARY 2012SUNRISE MEDICALSection 9Reference: Inhibit scheme

Page 81

QM-7 SERIES SERVICE MANUALPAGE 9.4JANUARY 2012 SUNRISE MEDICALSection 9Reference: R-NET Joystick & Omni Display Symbols

Page 82

QM-7 SERIES SERVICE MANUALPAGE 9.5 JANUARY 2012SUNRISE MEDICALSection 9Reference: Seat Height Adjustment MatrixSTFH (in)Front/Rear Post PositionPack

Page 83

QM-7 SERIES SERVICE MANUALPAGE 1.5 JANUARY 2012SUNRISE MEDICAL• Remove personal items of jewelry, such as rings, watches, chains etc. before workin