PULSE 6 SERVICE MANUAL
PAGE 5.7
SEPT 2011SUNRISE MEDICAL
Tilt Will Not Operate
1. Select tilt on the hand control (fi gure. 5.7.1).
Operate the joystick and listen closely to the
control module for a click. This click indicates
that the control module is closing the power relay
inside and supplying power to the tilt actuator.
If no click is heard, replace the control module.
Refer to Control Module Replacement section this
manual. Retest as necessary. If click is heard
proceed to step 3.
2. Verify that tilt is selected on the hand control.
Operate the joystick and listen closely to the tilt
actuator. If the actuator appears to be running
but the tilt is not moving, replace the tilt actuator.
Refer to Tilt Actuator and Micro-Switch Removal
section of this manual. Retest as necessary.
3. Locate the 6-pin connector on the rear of the tilt
See "B" (fi gure 5.6.2 previous page).
4. Disconnect the 6-pin cable leading to the
control module.
The tilt actuator used in the tilt system receives
power through a 6-pin connector located at the
rear of the tilt See "B" (Figure 5.6.2). The actuator
also contains micro-switches that open at the
end of their stroke to prevent stressing of the tilt
system. Around these micro switches are diodes
that allow reverse power to fl ow to the actuator
when reversing direction.
5. Select tilt on the hand control and place a
rubber band around the joystick to hold it displaced
(fi gure 5.7.2).
R-net Fault Codes cont.
Section 5
fig. 5.7.1
fig. 5.7.2